Scientific Equipments, Instruments for Plant Science and Environmental Science Research
 
New Products at Nu-tech International
 
All Products - Scientific Equipments, Instruments For Plant Science, Environmental Research
PP Systems
Hansatech Instruments Ltd.
Photon Systems Instruments
Decagon Devices, Inc.
PMS Instrument Company
Spectrum Technologies, Inc.
Tetracam Inc.
NTech Industries, Inc.
Odyssey Data Recording Systems - Dataflow Systems Pty Limited
Environmental Measuring Systems Brno
Rimik Pty Ltd. (Agridry)
UMS GmbH Munich
Yeo-Kal Electronics Pty. Ltd.
 
 

Product Specifications

PP Systems - Soil Respiration System: A portable system based on the EGM

   
The Soil Respiration System consists of the SRC-1 Soil Respiration Chamber and the EGM Environmental Gas Monitor. A chamber of known volume is placed on the soil and the rate of increase in CO2 within the chamber is monitored. The air is continuously sampled in a closed circuit through the EGM , and the soil respiration rate is calculated, displayed and recorded by the analyser. The air within the chamber is carefully mixed to ensure representative sampling without generating pressure differences which would affect the evolution of CO2 from the soil surface.

The system can be readily adapted to animal respiration studies or to other measurements of CO2 gas exchange.

  • Choice of two systems available (for EGM or CIRAS)
  • Simultaneous measurement of soil temperature as an optional extra
  • Robust chamber construction
  • Ergonomic system design allows for rapid and easy measurements
  • On-line statistical data analysis
  • Full data storage capability
  • RS232 output for transfer to a computer or printer
  • Low cost (EGM based system)

Once the SRC-1 chamber has been placed on the soil, the rate of increase in CO2 in the system should be linear, though any exchange with the outside air will cause it to decline with time. The CO2 concentration is measured every 8 seconds, and a quadratic equation fitted to the relationship between the increasing CO2 concentration and elapsed time. The coefficients of this are used to determine the rate of increase at time 0. A warning is given if the relationship is excessively non-linear.

 

Technical Specifications

  • Inputs
    Soil Temperature: 0 - 1 V = 0-50°C
  • Storage
    Capacity to store up to 1,500 records
  • Record Options
    Automatic logging after 120 s, or if the system CO2 concentration
    increases by more than 60 ppm Manual logging by key press
  • Power Supply
    12 V 2.0 Ah lead acid battery giving up to 5 hours continuous use.
  • Range
    0 - 9.99 g CO2 m-² hour-¹
  • Fan
    12 V DC
  • Housing
    PVC and stainless steel
  • Dimensions
    100 mm Ø x 150 mm H (excluding handle)
  • Weight
    900 g

More about SRS1

     
     
Home  |  All Products  |  New Products  |  About Us  |  Our Clients  |  Send Query  |  Support  |  Job Opportunities